The Indiana State Department of Health regulates the use of all x-ray machines. A full list of laws and regulations can be found at: http://www.in.gov/isdh/24351.htm
A physicist or x-ray machine inspector must evaluate x-ray equipment and programs at regular intervals. Dental x-ray facilities once every 3 years, podiatric and veterinary every 2 years, and all others every year.
X-Rays in the Healing Arts: Diagnostic x-ray systems 410 IAC 5-6.1 Sec. 119. (a)
- Podiatric and veterinary x-ray facilities shall be evaluated at least once each twenty-four (24) months by a diagnostic imaging physicist or x-ray machine inspector approved by the department.
- Dental x-ray facilities shall be evaluated at least once each thirty-six (36) months.
- All other diagnostic x-ray systems (such as medical and chiropractic) shall be evaluated at least once each twelve (12) months.
Before installation of new x-ray machines, a shielding assessment must be done. An accurate shielding assessment will determine how much and if lead is necessary in walls. After installation, a safety survey to check for scatter radiation in the surrounding areas needs to be performed.
410 IAC 5-6.1-118 bb:
- Floor plans and equipment arrangements for all new diagnostic x-ray installations, or modifications of such installations, shall be evaluated with regard to shielding requirements prior to construction. A safety survey shall be performed prior to first use. Both evaluations shall be performed by a diagnostic imaging physicist or a health physicist approved by the department.
Non-Medical (industrial/commercial) 410 IAC 5-5-3.2 (3)
Be evaluated at intervals not to exceed 1 year to assure compliance with the applicable requirements as specified in 410 IAC 5-5-3.2(a)(1).