The Indiana State Department of Health regulates the use of all x-ray machines.  A full list of laws and regulations can be found at:  http://www.in.gov/isdh/24351.htm

 

A physicist or x-ray machine inspector must evaluate x-ray equipment and programs at regular intervals.  Dental x-ray facilities once every 3 years, podiatric and veterinary every 2 years, and all others every year.

X-Rays in the Healing Arts:  Diagnostic x-ray systems  410 IAC 5-6.1 Sec. 119. (a)

  • Podiatric and veterinary x-ray facilities shall be evaluated at least once each twenty-four (24) months by a diagnostic imaging physicist or x-ray machine inspector approved by the department.
  • Dental x-ray facilities shall be evaluated at least once each thirty-six (36) months.
  • All other diagnostic x-ray systems (such as medical and chiropractic) shall be evaluated at least once each twelve (12) months.

 

Before installation of new x-ray machines, a shielding assessment must be done.  An accurate shielding assessment will determine how much and if lead is necessary in walls.  After installation, a safety survey to check for scatter radiation in the surrounding areas needs to be performed.

410 IAC 5-6.1-118 bb:

  • Floor plans and equipment arrangements for all new diagnostic x-ray installations, or modifications of such installations, shall be evaluated with regard to shielding requirements prior to construction. A safety survey shall be performed prior to first use. Both evaluations shall be performed by a diagnostic imaging physicist or a health physicist approved by the department.

 

 

Non-Medical (industrial/commercial)  410 IAC 5-5-3.2 (3)

Be evaluated at intervals not to exceed 1 year to assure compliance with the applicable requirements as specified in 410 IAC 5-5-3.2(a)(1).